GM-10
Detector
A low voltage interface connects the RAM DA 2000 monitor with the probe.The GM-10 uses a pancake GM detector with a thin mica window. This is encased in a rugged, splash proof housing for protection against shock, vibration and humidity. The probe also contains a high voltage power supply, pulse shaper, GM saturation indicator, detector identifier and malfunction detection circuitry. This configuration minimizes noise, improving sensitivity and stability.
Specifications
|
GM Type: |
|
Pancake LND 73118 or equivalent |
|
Window: |
|
Mica 1.5 - 2.0 mg/cm2 protected by a fine stainless steel wire mesh |
|
Sensitivity (137Cs): |
|
Approximately 350cpm/µSv/hr (3500cpm/mR/hr) |
|
Accuracy: |
|
±15% of reading |
|
Count Rate Range: |
|
0 - 42kcps |
|
Output Signals: |
|
TTL pulses. |
|
Temperature Range: |
|
Operation: -10oC to +50oC (15oF - 122oF) |
|
Humidity Range: |
|
40% to 95% RH (non condensing) |
|
Dimensions: |
|
260mm (10.2") length, 70mm (2.8") width, 75mm (2.9") height |
|
Weight: |
|
370g (0.8lbs) |
|
Casing: |
|
Aluminum, splash proof |
|
Performance (surface sensitivity - in contact) Beta Emitting Isotopes |
||||
|
Isotope |
Emax (keV) |
2p Effic (%) |
Sensitivity cpm/Bq/cm2 |
MDL* Bq/cm2 |
|
14C |
156 |
3 |
15 |
5 |
|
35S |
167 |
8 |
40 |
2 |
|
147Pm |
224 |
15 |
75 |
1 |
|
99Tc |
290 |
20 |
100 |
0.8 |
|
90Sr + 90Y |
580 + 2280 |
50 |
500 |
0.2 |
|
67GA |
900 |
48 |
235 |
0.35 |
|
123I |
1000 |
55 |
280 |
0.3 |
|
36Cl |
714 |
45 |
210 |
0.4 |
|
210Pb |
1160 |
50 |
250 |
0.3 |
|
32P |
1710 |
70 |
330 |
0.3 |
|
51CR |
325 |
0.2 |
1 |
75 |
|
99mTc |
140 |
5.3 |
25 |
3 |
|
125I |
35 |
0.5 |
2.4 |
32 |
|
131I |
610 |
40 |
190 |
0.4 |
|
230Th |
4600 |
14 |
65 |
1.1 |
|
241Am |
5400 |
14 |
65 |
1.1 |
*MDL - Minimum detectable level calculations are based on background readings of 1 cps and a confidence level of 99%
Ordering Information
GM-10 - 2" Pancake BAK-1170
Note: Specifications subject to change without notice.