PM-10
Detector
A low voltage interface connects the RAM DA 2000 monitor with the probe.The PM-10 contains a highly efficient, hermetically sealed, NaI(Tl) scintillator, 2" diameter and 0.04" thick, coupled to a 2" photomultiplier tube. This is encased in a rugged, splash proof housing for protection against shock, vibration and humidity. The probe also contains a high voltage power supply, single channel analyzer (SCA), pulse shaper, GM saturation indicator, detector identifier and malfunction detection circuitry. This configuration minimizes noise, improving sensitivity and stability.
Specifications
|
Radiation Detected: |
|
Low energy gamma, 10keV to 80keV |
|
Scintillator: |
|
NaI(Tl) 2" diameter, 0.04" thick. Window of 1 mil. (0.001") aluminum |
|
Energy Calibration: |
|
SCA within the energy range ±5keV (minimum) |
|
Count Rate Range: |
|
0 - 50,000cps |
|
Output Signals: |
|
TTL pulses. |
|
Temperature Range: |
|
Operation: -10oC to +50oC (15oF - 122oF) |
|
Humidity Range: |
|
40% to 95% RH (non condensing) |
|
Dimensions: |
|
290mm (11.4") length, 70mm (2.8") width |
|
Weight: |
|
1.25kg (2.8lbs) |
|
Casing: |
|
Aluminum, splash proof |
Surface Sensitivity (in contact):
|
Isotope |
Sensitivity cpm/Bq/cm2 |
MDL Bq/cm2 |
|
51Cr |
4.60 / 0.65 |
150 / 225 |
|
125I |
460 / 440 |
1.50 / 0.35 |
|
131I |
90 / 35 |
7.50 / 4.30 |
|
99mTc |
150 / 30 |
4.50 / 4.50 |
|
90Sr+90Y |
280 / 40 |
2.50 / 3.50 |
|
36Cl |
8.0 / 3.50 |
85 / 43 |
|
210Pb |
95 / 35 |
7.40 / 4.30 |
|
32P |
180 / 30 |
4.0 / 4.50 |
(a) without SCA calibration (open window)
(b) Energy calibrated to 125I (approx. 30keV)
* Minimum detectable level calculations are based on background reading of 600 and 120 cpm for the detector with an open window (no energy calibration) and the detector calibrated to 125I respectively. The confidence level is 99%
Ordering Information
PM-10 - 2" x 0.04" Scintillator (Open Window) BAK-1490
Note: Specifications subject to change without notice.