Home page › PM-10

PM-10 Detector

The PM-10 is a gamma scintillation probe for detection of low energy gamma and X-rays. It can be purchased optimized for detection of 125I and other low energy gamma emitters. The combination of the RAM DA 2000 monitor and the high efficiency of the PM-10 provides a minimum detectable level (MDL) of 0.35Bq/cm2 (approximately 1.10-5 µCi/cm2 surface contamination of 125I.

A low voltage interface connects the RAM DA 2000 monitor with the probe.The PM-10 contains a highly efficient, hermetically sealed, NaI(Tl) scintillator, 2" diameter and 0.04" thick, coupled to a 2" photomultiplier tube. This is encased in a rugged, splash proof housing for protection against shock, vibration and humidity. The probe also contains a high voltage power supply, single channel analyzer (SCA), pulse shaper, GM saturation indicator, detector identifier and malfunction detection circuitry. This configuration minimizes noise, improving sensitivity and stability.

Specifications

Radiation Detected:

 

Low energy gamma, 10keV to 80keV

Scintillator:

 

NaI(Tl) 2" diameter, 0.04" thick. Window of 1 mil. (0.001") aluminum

Energy Calibration:

 

SCA within the energy range ±5keV (minimum)

Count Rate Range:

 

0 - 50,000cps

Output Signals:

 

TTL pulses.
Detector status: ID, OK, malfunction, overflow

Temperature Range:

 

Operation: -10oC to +50oC (15oF - 122oF)
Storage: -20oC to +60oC (-5oF - 140oF)

Humidity Range:

 

40% to 95% RH (non condensing)

Dimensions:

 

290mm (11.4") length, 70mm (2.8") width

Weight:

 

1.25kg (2.8lbs)

Casing:

 

Aluminum, splash proof

Surface Sensitivity (in contact):

Isotope

Sensitivity cpm/Bq/cm2
(a) / (b)

MDL Bq/cm2
(a) / (b)

51Cr

4.60 / 0.65

150 / 225

125I

460 / 440

1.50 / 0.35

131I

90 / 35

7.50 / 4.30

99mTc

150 / 30

4.50 / 4.50

90Sr+90Y

280 / 40

2.50 / 3.50

36Cl

8.0 / 3.50

85 / 43

210Pb

95 / 35

7.40 / 4.30

32P

180 / 30

4.0 / 4.50

(a) without SCA calibration (open window)
(b) Energy calibrated to 125I (approx. 30keV)

* Minimum detectable level calculations are based on background reading of 600 and 120 cpm for the detector with an open window (no energy calibration) and the detector calibrated to 125I respectively. The confidence level is 99%

Ordering Information

PM-10 - 2" x 0.04" Scintillator (Open Window) BAK-1490

Note: Specifications subject to change without notice.