PM-11
Detector
A low voltage interface connects the RAM DA 2000 monitor with the probe.The PM-11 contains a highly efficient, hermetically sealed, NaI(Tl) scintillator, 2" diameter and 2" thick, coupled to a 2" photomultiplier tube. This is encased in a rugged, splash proof housing for protection against shock, vibration and humidity. The probe also contains a high voltage power supply, single channel analyzer (SCA), pulse shaper, GM saturation indicator, detector identifier and malfunction detection circuitry. This configuration minimizes noise, improving sensitivity and stability.
Specifications
|
Radiation Detected: |
|
Gamma above 50 keV |
|
Scintillator: |
|
NaI(Tl) 2" diameter, 2" thick. Window of 1 mm. (0.04") aluminum |
|
Energy Calibration: |
|
Optional factory calibrated single channel analyzer (SCA) within the energy range |
|
Count Rate Range: |
|
0 - 50,000cps |
|
Output Signals: |
|
TTL pulses. |
|
Temperature Range: |
|
Operation: -10oC to +50oC (15oF - 122oF) |
|
Humidity Range: |
|
40% to 95% RH (non condensing) |
|
Dimensions: |
|
340mm (13.4") length, 70mm (2.75") width |
|
Weight: |
|
1.75kg (3.9lbs) |
|
Casing: |
|
Aluminum, splash proof |
Surface Sensitivity (in contact)
|
Isotope |
Sensitivity cpm/Bq/cm2 |
MDL Bq/cm2 |
|
51Cr |
27 |
140 |
|
125I |
25 |
150 |
|
131I |
320 |
12 |
|
99mTc |
315 |
12 |
|
32P |
17 |
220 |
* Minimum detectable level calculations are based on background reading of 3600 cpm The confidence level is 99%
Ordering Information
PM-11- 2" x 2 Scintillator (Open Window) BAK-1500
Note: Specifications subject to change without notice.