Production Monitoring
The module is based on five detectors and an electronic board with two 0-10V analog outputs. The electronic board is used to amplify the current, received from a PN photodiode coupled to a scintillation crystal.
|
Detector specifications |
|
|
Type |
SD100 |
|
Radiation detection |
Gamma above 100 keV |
|
Detector |
PN photo diode coupled with CsI (Tl) Scintillated |
|
Sensitivity |
2.6mV / 1mR/h |
|
Measuring Range |
0.01R/h – 8000R/h* |
|
Temperature Range |
Operation: -10° C to + 50° C |
|
Dimensions |
35 mm long x 23 mm diameter |
|
Connector |
BNC |
* - 1 R/h is equivalent to the radiation field of 1 Ci of FDG at a distance of 22mm
Features and Specifications
Input voltage ± 15V
Offset adjustment
Linearity ± 10% in the range of 1 ¸ 8000R/h
Up to 5 meters cable between the detectors and the electronics
Two analog outputs voltage 0 - 10V for logging and control
On site calibration of the 0 - 10V output voltage for 0 - 5Ci activity range,
by gain adjustment.
Logic selection between detectors
Optional local display