RAM SURF

The RAM SURF uses the same detector housed in an external probe which offers the extra flexibility required for monitoring less accessible positions.
Features
Large, four digit, continuous ranging, LCD display
Simple, three key operation
Large 1.75" pancake GM detector
Suitable for alpha, beta or gamma detection
Readout of count rate
Units of either counts per minute or counts per second
Programmable alarm thresholds
Audible and visual (LED) count rate indication
Smart software provides fast, yet stable response
Dead time correction
Alarms for overflow, low battery and detector failure
Automatic self diagnostic routines
Durable case with splash resistant keypad
Specifications
|
Display: |
|
Large, four digit, continuous ranging, LCD readout for diplay of accumulated dose, dose rate, alarm threshold value, overflow, low battery, and detector failure |
|
LED: |
|
Flashing count rate indication |
|
Audio: |
|
Piezo electric element synchronized with count LED |
|
Units: |
|
Counts per second (cps) or counts per minute (cpm) |
|
Controls: |
|
Three key keypad with positive feedback for ON/OFF, RESET and SPEAKER. Key combinations for selecting threshold displays |
|
Power Source: |
|
Standard 9V alkaline battery - provides approximately 50 hours continuous use with speaker off. Automatic battery check under full load. |
|
Detector: |
|
1.75" Pancake GM tube - LND73118 or equivalent |
|
Window: |
|
Mica, 1.5 - 2.0 mg/cm2 protected by stainless steel mesh |
|
Count Range: |
|
0 - 42,000 cps |
|
Sensitivity: |
|
Approximately 5.8 cps/µSv/hr (58 cps/mR/hr) - 137Cs |
|
Background: |
|
0 - 1 cps |
|
Temperature Range: |
|
Operation: -10oC to +50oC (15oF to 122oF) |
|
Humidity Range: |
|
40% to 95% RH (non condensing) |
|
Case: |
|
High impact ABS |
|
Dimensions: |
|
7.2cm x 11.4cm x 6.2cm (2.8" x 4.5" x 2.4") |
|
Weight: |
|
Gene: 0.340 Kg (0.75 lbs) Surf: 0.630 Kg (1.4 lbs) |
|
Performance (surface sensitivity - in contact) Beta Emitting Isotopes |
||||
|
Isotope |
Emax (keV) |
2p Efficency(%) |
Sensitivity cps/Bq/cm2 |
MDLBq/cm2 |
|
14C |
156 |
3 |
0.25 |
9.0 |
|
35S |
167 |
8 |
0.65 |
3.7 |
|
147Pm |
224 |
15 |
1.25 |
2.0 |
|
99Tc |
290 |
20 |
1.75 |
1.5 |
|
90Sr + 90Y |
580 + 2280 |
50 |
8.50 |
0.3 |
|
36Cl |
714 |
45 |
3.50 |
0.7 |
|
210Pb |
1160 |
50 |
4.00 |
0.6 |
|
32P |
1710 |
70 |
5.50 |
0.5 |
|
Additional radionuclides |
||
|
Isotope |
Sensitivity cps/Bq/cm2 |
MDL Bq/cm2 |
|
99mTc |
0.45 |
5.5 |
|
125I |
0.04 |
63.0 |
|
131I |
3.00 |
0.8 |
|
230Th |
1.10 |
2.5 |
|
241Am |
1.10 |
2.5 |
*MDL - Minimum detectable level calculations are based on background readings of 1 cps and a confidence level of 99%
Ordering Information
RAM SURF-CPS (Telephone Connector) BAK-2080
RAM SURF-CPM (Telephone Connector) BAK-2090
Note: Specifications subject to change without notice.